Click here to purchase

1.1 The scope given in Clause 1 of IEC 601-1 applies except that 1.1 shall be replaced by the following:

1.1.1 This specification includes requirements for the safety and performance of pulse oximeters, as defined in 3.8 intended for use in approximating the saturation of arterial hemoglobin, non-invasively, from a light signal transmitted through the tissue, taking into account the pulsatile nature of volume changes with blood flow.

1.2 The field of application includes, but is not limited to: intraoperative and perioperative use; adult critical care application; pediatric and neonatal application; and general determination of saturation on hospitalized and non-hospitalized patients.

1.3 Pulse oximeters intended for use in laboratory research applications and “bench” type oximeters that require a blood sample from the patient are outside the scope of this specification.

Product Details

Published:
01/01/2000
Number of Pages:
7
File Size:
1 file , 50 KB