1.1 This practice covers general procedures necessary for the calibration of magnification of scanning electron microscopes. The relationship between true magnification and indicated magnification is a complicated function of operating conditions.2 Therefore, this practice must be applied to each set of standard operating conditions to be used.
Product Details
- Published:
- 01/01/2014
- Number of Pages:
- 6
- File Size:
- 1 file , 200 KB