Click here to purchase

1.1 This practice covers procedures for detecting the sources of gas leaking at the rate of 1 × 105 Pa m³/s (1 × 104 standard cm³/s) or greater. The tests may be conducted on any object that can be pressurized with a tracer gas that is detectable by a thermal conductivity detector. The test sensitivity will vary widely depending on the tracer gas used.

1.2 UnitsThe values stated in either SI or std-cc/sec units are to be regarded separately as standard. The values stated in each system may not be exact equivalents: therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Product Details

Published:
07/01/2011
Number of Pages:
3
File Size:
1 file , 84 KB
Redline File Size:
2 files , 170 KB