Click here to purchase

1.1 This practice provides guidelines for calibration and instrumentation (both analog and computerized systems) used in the psychophysiological detection of deception (verification of truth). As a minimum, such instrumentation shall simultaneously record an individual’s respiratory, electrodermal, and cardiovascular activity. Analog polygraphs shall be calibrated by the psychophysiological detection of deception (PDD) examiner, manufacturer, or factory-authorized individual. Computerized instrumentation shall be calibrated by the manufacturer or factory-authorized individual. The PDD examiner or factory-authorized individual shall perform functionality checks to ensure instrumentation is operating properly.

1.2 This practice does not prohibit additional components which may be offered as supplemental measurements of physiological change. Additional recording components (such as movement sensors) shall meet the manufacturer’s specifications.