Click here to purchase

1.1 This test method covers the determination of sulfur in organic compounds by sodium peroxide bomb ignition. It is intended for application to samples of organic materials containing more than 0.5% sulfur. The presence of iron may require a separation such as is described.

1.2 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use . For specific hazard statements, see Section 7 and Notes 2 and 3.

Note 1-A test method based on an oxygen bomb combustion of petroleum products is described in Test Method D129.

Product Details

Published:
09/10/1998
Number of Pages:
4
File Size:
1 file , 39 KB