1.1 In harmony with the Joint Committee for Guides in Metrology (JCGM) and detection concepts of the International Union of Pure and Applied Chemistry (IUPAC) (1, 2)2, this test method uses a series of replicated measurements of an analyte at dosage levels giving instrumental responses that bracket the critical value, a truncated normal distribution model, and confidence bounds to establish a standard for estimating practical and statistically robust limits of detection.
Note 1: Other standards are available that evaluate the general performance of detection technologies for various analytes in complex matrices (for example, Practice E2520).
Product Details
- Published:
- 02/01/2020
- Number of Pages:
- 8
- File Size:
- 1 file , 110 KB
- Redline File Size:
- 2 files , 230 KB