1.1”This practice covers procedures for use in the calibration of volumetric instruments that include glassware, plasticware, and laboratory standards that are in common use in chemical, analytical, clinical, and calibration laboratories. It is based on the gravimetric determination of the quantity of pure water, either contained or delivered at a calibration temperature, and the conversion of this value to a volume at a given reference temperature, normally 20””’C by means of suitable equations. Calibration using mercury is excluded. Calibration may be performed using alternative gravimetric methodology, if it is demonstrated and documented that the results obtained are equivalent to those obtained using the methodology described herein. Alternative reference temperatures and associated equations are provided.